Recall Z-2655-2023 ยท DSJ Alarm, Blood-Pressure

Philips North America Llc

EarlyVue VS30 Vital Signs Monitor, utilizing E01 (Oridion Microstream sidestream etCO2) and R21 (Masimo Rainbow RRa) simultaneously, Product Code 863359

DateSep 29, 2023 (initiated Aug 31, 2023)
StatusOpen, Classified
ReasonA software issue prevents the monitor from alarming when the high/low acoustic respiration rate (RRa) limits are violated.
Root causeSoftware design
ActionAn URGENT Medical Device Correction notification letter dated 8/29/23 was sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users If you use EarlyVue VS30 devices with both options E01 (Oridion Microstream Sidestream etCO2) and option R21 (Masimo Rainbow RRa): 1) Observe the EarlyVue VS30 monitor display screen for out-of-limit acoustic respiration rates (RRa) during procedures when there is simultaneous use of E01 (Oridion Microstream sidestream etCO2) and R21 (Masimo Rainbow RRa), while Respiration Source Mode is set to Auto, an
Quantity13 total devices
DistributionWorldwide - US Nationwide distribution in the states of MA, MI. The country of Brazil.
FDA recordOpen on fda.gov

510(k) clearances named in this recall