Recall Z-0093-2024 ยท BRY Cabinet, Table And Tray, Anesthesia

CareFusion 303, Inc.

BD Pyxis CII Safe ES, REF: 1116-00; and BD Pyxis : CII Safe ES Desktop PC, Wired SCNR, REF: 107-254-01; CII Safe ES Desktop PC, W-LESS SCNR, BIO, REF: 107-255-01; Enterprise Server, REF: 1115-00, 134157-01, 137670-02, 134148-01, 135292-01, 136449-01, 133760-01

DateOct 10, 2023 (initiated Sep 6, 2023)
StatusOpen, Classified
ReasonWhen global edit is used to update multiple formulary properties simultaneously, the following properties: 1) Require Lot Number on Recall, 2) Access Destruction Bin (Witness), 3) Add to Destruction Bin (Count/Empty), 4) CII Safe Stock Out Notice are, without user warning, overwritten to the value NO , which could result in diversion of controlled substances from the automated dispensing cabinet.
Root causeSoftware change control
ActionOn 9/6/23, correction notices were mailed to customers informing them to do the following: 1) Do not use the global edit feature until your software is updated. 2) If a global edit has already been performed, it is recommended to verify that the formulary device settings are accurate for the 4 configuration settings. Controlled substances that have been edited via the global edit feature should be reviewed retrospectively for discrepancies in the destruction bin and a manual witness record may be needed until the settings are adjusted for both "Access Destruction Bin" and "Add to Destruction
Quantity61
DistributionUS Nationwide distribution in the states of LA, NC, NY, MI, MA, NV, KS, IA, CA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.