Recall Z-2631-2023 ยท MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator

Philips Respironics, Inc.

Philips Respironics V60/V60 Plus Ventilator, PCBA Part Number: 453561544461 V60 Device Part Numbers 1053617 (US) and 1137276 (Philippines)

DateSep 28, 2023 (initiated Aug 25, 2023)
StatusOpen, Classified
ReasonPower Management PCBAs may malfunction, causing a power failure leading to ventilator loss of function.
Root causeNonconforming Material/Component
ActionImpacted customers were contacted via phone and email on August 25, 2023. Philips released an "Urgent Medical Device Correction" customer letter to make customers aware of the potential for a power failure leading to ventilator loss of function. Instructions to Customer Discontinue use of and quarantine these device(s) until you are contacted by a Philips Respironics representative to remove and/or replace the Power Management PCBAs in the affected device(s). Philips Respironics will contact you to coordinate a time to remove and/or replace the impacted component located within your device(s
Quantity15 (14 US; 1 OUS)
DistributionUS Distribution to states of: OH, CA, SC, OR; and OUS country of: Philippines.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102985V60 VENTILATOR WITH PPV AND AUTO-TRAK+ SOFTWARE OPTIONS
Respironics, Inc., Apr 5, 2011