ClearanceBrief
Recall Z-0029-2024 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Fresenius Medical Care Holdings, Inc.

2008T Hemodialysis System without CDX

DateOct 6, 2023 (initiated Sep 6, 2023)
StatusOpen, Classified
ReasonPotential PCBA leaching from tubing of hemodialysis machines.
Root causeDevice Design
ActionConsignees were sent an URGENT MEDICAL DEVICE CORRECTION NOTICE via USPS, dated 09/06/2023. The notice informed consignees that there is a potential for NDL-PCBAs could leach from silicone tubing manufactured with a peroxide-based catalyst. The firm also states that it has been demonstrated that dialysate run through 2008 Series machines used during typical maintenance HD treatments for at least 36 days, cumulatively, showed levels of NDL-PCBAs that were below toxicologically relevant thresholds. Consignees are to contact FMCRTG Technical Service at 1-800-227-2572 to schedule an update to devi
Quantity733 units
DistributionDomestic: Nationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K1507082008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T Hemodialysis Machine
Fresenius Medical Care Renal Therapies Group, LLC, Dec 10, 2015
K1739722008T BlueStar Hemodialysis Machine
Fresenius Medical Care Renal Therapies Group, LLC, May 10, 2018
K2229522008T BlueStar Hemodailysis Machine
Fresenius Medical Care Renal Therapies Group, LLC, Oct 27, 2022