Recall Z-0036-2024 ยท FQO Table, Operating-Room, Ac-Powered

Getinge Usa Sales Inc

Meera Mobile Operating Table-to support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment. Model Number: 7200.01F2

DateOct 4, 2023 (initiated Sep 6, 2023)
StatusOpen, Classified
ReasonUnder certain conditions, an issue might prevent the device from performing as intended. A specific sequence of commands on the control unit activates the traction drive and triggers an unintended driving (autodrive) of max. 7 sec. may result in the following reasonably foreseeable injury or damage to the health of patients and/or users
Root causeSoftware Manufacturing/Software Deployment
ActionMaquet GMBH / Getinge initiated Urgent Medical Device Correction letter on 9/6/23 via Fed'X. Letter states reason for recall, health risk and action to take: Please examine your inventory immediately to determine if you have any of the Meera Mobile Operating Table with the serial numbers listed in this notice. Maintain awareness on this notice and related actions until your operating table has been updated to ensure effectiveness of the corrective action. Please complete and sign the attached MEDICAL DEVICE CORRECTION- RESPONSE FORM (page 4) to acknowledge that you have received this notificat
Quantity224 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.