Recall Z-2520-2023 ยท JAK System, X-Ray, Tomography, Computed

Philips Healthcare

Incisive CT, software version 5.0

DateSep 30, 2023 (initiated Aug 2, 2023)
StatusOpen, Classified
ReasonPhilips has identified three software issues with compliance concerns to Subchapter J: Issue 1: Coronary Computed Tomography Angiography (CTA) scan may not be triggered at the optimal time. Issue 2: Use of aborted surview for scan planning will result in deviation from intended scan area. Issue 3: Actual clinical scan with bariatric (Noah) couch will deviate from the intended scan area if surview scan length is greater than 1832mm.
Root causeRadiation Control for Health and Safety Act
ActionPhilips will inform customers of the issue via an Electronic Product Radiation Defect notification. Philips will schedule an appointment with customers to install the software update.
Quantity140 systems
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov