Recall Z-0027-2024 ยท KNW Instrument, Biopsy

Bard Peripheral Vascular Inc

Bard Marquee Disposable Core Biopsy Instrument Kit, REF: MQK2016

DateOct 4, 2023 (initiated Aug 10, 2023)
StatusOpen, Classified
ReasonDiameter mismatch between coaxial and biopsy needle (cutting cannula) in disposable core biopsy instrument kits, where the coaxial cannula inner diameter prohibits the biopsy needle (cutting cannula) from properly fitting into the coaxial cannula and accessing target tissue, which may lead to repeated procedure, and procedural complications, such as pain/discomfort, pneumothorax, and bleeding.
Root causeDevice Design
ActionOn 8/10/23, BD mailed or emailed recall notices to Medical Directors, Risk Managers, Medical Device Safety Offices, Lab Managers, Clinical Engineers, and Distributors who were told to do the following: 1) Discontinue distribution and use and immediately dispose of all affected product. 2) Share this notification with all users within your facility network of the product to ensure they are also aware of this field action. 3) Distributors, identify all customers within your distribution network that purchased any affected product and provide a copy of the customer letter to all customers and a
Quantity1,545
DistributionWorldwide - US Nationwide distribution in the states of PR, CO, TX, CA, WA, OR, AZ, NJ, NY, OH, NC, MD, MI, LA, IN, AK, TN and the countries of CA, AU, BE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133948BARD(R) MONOPTY(R) DISPOSABLE CORE BIOPSY INSTRUMENT, BARD(R) MAX-CORE(R) DISPOSABLE CORE INSTRUMENT
Bard Peripheral Vascular, Inc., Feb 21, 2014