Recall Z-2665-2023 ยท OWD Somatic Gene Mutation Detection System

Qiagen GmbH

therascreen KRAS RGQ PCR Kit (24)-IVD qualitative PCR assay used on the Rotor-Gene Q MDx instrument for the detection of seven somatic mutations in the human KRAS oncogene REF 870021 (US IVD)

DateSep 29, 2023 (initiated Aug 22, 2023)
StatusOpen, Classified
ReasonHandbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_INT_CTRL_FAIL. A dilution of CRC tissue could lead to a false negative result. To remove ambiguity, an alteration must be made to the S the sample result with this flag should be interpreted as invalid
Root causeUnder Investigation by firm
ActionQiagen US issued Urgent Medical Device Correction Letter to US consignees -therascreen KRAS RGQ PCR Kit (24), REF 870021 on 8/22/23 via email. Letter states reason for recall, health risk and action to take: This notice only applies to you if you downloaded and used the current version (Revision 13, January 2023) of the Instructions for Use. If you downloaded and used the current version (Revision 13, January 2023) of the Instructions for Use as mentioned above, follow these actions: o The dilution step of CRC tissue sample is incorrectly indicated and should not be followed. Follow the corre
DistributionWorldwide - US Nationwide distribution in the states of AZ, CA, FL, IL, IN, KS, MA, NC,NY,OR, PA, TN, TX and the countries of KR, CN
FDA recordOpen on fda.gov