ClearanceBrief
Recall Z-0124-2024 ยท FRN Pump, Infusion

B. Braun Medical, Inc.

8713052U, Infusomat Large Volume Pump, Non-Wireless BATTERY PACK; Volumetric Infusion Pump System

DateOct 25, 2023 (initiated Sep 21, 2023)
StatusOpen, Classified
ReasonUpstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to hemodynamic instability requiring medical intervention to prevent permanent impairment to body structures or body functions. In some cases, this may be life-threatening or may lead to death.
Root causeComponent design/selection
ActionOn September 21, 2023, the firm began notifying customers of the issue via "URGENT MEDICAL DEVICE CORRECTION" letters. BBMI will schedule service to update impacted devices. Customers were provided with stickers to identify affected units while waiting for the service to be completed. In the meantime, customers were informed of measures that may aid in continued use of the pumps, such as only using non-impacted devices for the delivery of high-risk medications and placing the impacted units in care areas that do not administer high risk medications. Customers may contact BBMI by calling 1-8
Quantity884
DistributionDomestic distribution nationwide. Foreign distribution to Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142596Infusomat Space Volumetric Infusion Pump
B.Braun Medical, Inc., Jun 18, 2015
K191910SpaceStation MRI
B. Braun Melsugen AG, Mar 12, 2020