Recall Z-2635-2023 ยท FRN Pump, Infusion
Fresenius Kabi USA, LLC
Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.
| Date | Sep 25, 2023 (initiated Apr 22, 2022) |
| Status | Open, Classified |
| Reason | Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set |
| Root cause | Device Design |
| Action | Ivenix initiated contact on 4/22/22/ via phone communicating the contents of the customer letter which was then emailed to each customer. Letter states reason for recall, health risk and action to take: Please note, the specified start-up flow accuracy of up to +8% is maintained throughout the extended start-up time; and the steady-state flow accuracy of +l-SYo of the LVP is not affected with this condition. There is no corrective action you need to take at this time. We believe that this condition is low risk in that the affected flow rate range is limited, the start'up flow rate accuracy is |
| Quantity | 1344 pumps |
| Distribution | US Nationwide distribution in the states of NJ, WI. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K183311 | Ivenix Infusion System (IIS) Ivenix, Inc., Jun 7, 2019 |