Recall Z-0025-2024 ยท FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days

Becton Dickinson Infusion Therapy Systems Inc.

REF: 386862, BD Cathena Safety IV catheter with BD Multiguard Technology, 20 GA 1.00 IN, 1.1 x 25 mm 64 mL/min , Rx Only, Sterile EO

DateOct 3, 2023 (initiated Aug 29, 2023)
StatusOpen, Classified
ReasonThere is the potential that the safety shield may not properly engage on IV catheters. The needle can pull through the safety shield and disconnect from the needle safety shield leaving the contaminated needle point exposed.
Root causeProcess change control
ActionOn August 29, 2023, Becton Dickinson issued an "Urgent Medical Device Recall" notification via. FedEx. Becton Dickinson asked consignees to take the following actions: 1. Immediately review your inventory for Catalog Number 386862, Lot number 2238464. Destroy all unused product subject to the recall following your institution's process for destruction. 2. For indwelling catheters, no further action is needed. 3. If a defective product was used on a patient and the used needle was disposed of without the clinician suffering a needlestick injury, no further action is necessary to further pro
Quantity98,000 units
DistributionUS Nationwide distribution in the states of CA, FL, GA, MA, MN, MO, MS, NJ, OH, TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K220584BD Cathena Safety IV Catheter
Becton Dickinson Infusion Therapy Systems, Inc., Mar 31, 2022