Recall Z-2651-2023 ยท BSK Cuff, Tracheal Tube, Inflatable

Covidien

Puritan Bennett Cuff Pressure Manager, REF: 180-03

DateSep 29, 2023 (initiated Aug 22, 2023)
StatusOpen, Classified
ReasonInadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which prevents the device from being charged, and could lead to treatment delay.
Root causeComponent design/selection
ActionOn 8/22/23 Medtronic mailed recall notices to risk managers and directors of respiratory care informing them of the following: 1) Discontinue use, quarantine, and return affected devices to the firm. 2) Complete and return the customer confirmation form via email to it to rs.gmbfcamitg@medtronic.com. 3) Forward the notice to all those who need to be aware within your organization or to any organization where the potentially affected product with the specified serial numbers has been transferred or distributed. If you have any questions regarding this communication, contact customer service
Quantity408
DistributionUS: AL, NJ, MA, MI, NY, NC, CO, TX, FL, KY, UT, CA, AZ, KS, IL, TN. OUS: El Salvador, Sweden, Canary Islands
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K202874Puritan Bennett Cuff Pressure Manager
Covidien, LLC, Jan 9, 2021