Recall Z-2629-2023 ยท HIG Insufflator, Hysteroscopic
WOM World Of Medicine AG
Aquilex Fluid Control System-Intended to provide fluid distension of the uterus during diagnostic and operative hysteroscopies and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus. Models AQL-100PBS and AQL-100P
| Date | Jan 5, 2024 (initiated Jul 17, 2023) |
| Status | Open, Classified |
| Reason | The display of inflow volume can reach its limit of 30,000 ml during long procdeure and the inflow volume display will freeze at the maximum value while the deficit will start counting backwards until 0 ml is reached and may result in fluid overload |
| Root cause | Software design |
| Action | W.O.M. World of Medicine GmbH (WOM), notified the US/Canada distributor Hologic via E-mail Urgent Field Safety Notice (FSN) - MEDICAL DEVICE RECALL Aquilex Fluid Control System Pump - REF AQL-100PBS and AQL-100P WOM reference no.: 2023-0001 on 7/17/23. Hologic notified sales representatives and the end users (hospitals and medical facilities) in the US and Canada on 8/21/2023 via Telephone calls and letter. Letter states reason for recall, health risk and action to take: W.O.M. WORLD OF MEDICINE GmbH has developed a software version, which, among others, starts emitting visual and acous |
| Quantity | 1287 units |
| Distribution | Nationwide Foreign: Canada, Australia, New Zealand, Belgium, France, Germany, Ireland, Netherlands, Spain, Sweden, Norway, Switzerland, UK, Kuwait, Qatar, Turkey, UAE |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K172040 | Aquilex Fluid Control System AQL-100S W.O.M. World of Medicine GmbH, Feb 5, 2018 |