Recall Z-0075-2024 ยท HIF Insufflator, Laparoscopic

Olympus Corporation of the Americas

Olympus High-Flow Insufflation Unit, Model UHI-4. For laparoscopic surgery.

DateOct 18, 2023 (initiated Sep 22, 2023)
StatusOpen, Classified
ReasonThere have been reports of patients suffering arrhythmias, reported as short cardiac arrests, during surgical procedures where UHI-4s were used. These events may have been due to an over insufflation of the abdominal cavity resulting from use of the UHI-4 during the procedures.
Root causeUnder Investigation by firm
ActionAn URGENT: MEDICAL DEVICE CORRECTIVE ACTION notification letter dated 9/22/23 was sent to customers. Action steps to be taken by the end user: Our records indicate that your facility has purchased one or more of the Olympus UHI-4. Olympus requires you to take the following actions: 1. Carefully read the content of this Field Corrective Action Letter. 2. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this letter. 3. Complete the enclosed response form and return it to our recall partner, Sedgwick, via e-mail at olympus5578@sedgwick.com or by fax
Quantity3136 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110294HIGH FLOW INSUFFLATION UNIT UHI-4
Olympus Medical Systems Corporation, Nov 18, 2011
K122180HIGH FLOW INSUFFLANTION UNIT
Olympus Medical Systems Corporation, Nov 13, 2012