Recall Z-2625-2023 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Remote Diagnostic Technologies Ltd.
Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, and 00-1007, when Used With Tempus Pro Video Laryngoscope, REF 01-2044
| Date | Sep 21, 2023 (initiated Aug 24, 2023) |
| Status | Open, Classified |
| Reason | Full screen error message may occur on patient monitors with affected hardware version, either during or after laryngoscopes are unplugged from monitors, informing users a 60-100 second shutdown/restart are required. Error prevents: viewing graphical patient vital signs, sounding of pulse audio tone, visualization of patient airway used during intubation; may lead to treatment delay or hypoxia. |
| Root cause | Device Design |
| Action | On 8/24/23 Correction Notices were mailed to customers and distributors who were asked to do the following: 1) If monitor has Trizeps-7 Hardware, remove the laryngoscope from service with this monitor. Users must use an alternative laryngoscope not connected to the Tempus Pro Monitor to manage the patient s airway to avoid interruption in patient care. The Tempus Pro Monitor can remain in service if the Tempus Pro Video Laryngoscope is not connected to the monitor. 2) If monitor has Trizeps-6 Hardware, the laryngoscope can continue to be used with the monitor. 3) Post this Urgent Medical Devi |
| Quantity | 3,524 |
| Distribution | US:AZ, AL, CA, CO, DC, DE, FL , GA, HI, IA, ID, IL, IN , KS, KY, MA , MD , ME, MI, MN, MO, MS, MT, NC , ND, NE, NJ, NM , NV, NY, OH, OK, OR, PA, TN, TX, UT, VA , WA, WI, WV, WY. OUS: AE, AT, AU, BE, BN, BO, CA, CH, CZ, DE, DK, ES, FR, GB, HK, IE, IL, IS, IT, KY, LU, MY, NL, NO, OM, PT, RO, SA, SE, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K201746 | Tempus Pro Patient Monitor Remote Diagnostic Technologies Limited, Sep 18, 2020 |