Recall Z-2587-2023 ยท OWB Interventional Fluoroscopic X-Ray System

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Philips Allura Xper, Allura Centron, Allura CV, Azurion, UNIQ Clarity, and MultiDiagnost-Eleva systems

DateSep 14, 2023 (initiated Aug 15, 2023)
StatusOpen, Classified
ReasonA damaged foot switch may cause instances where no or only intermittent x-ray radiation initiation is possible.
Root causeDevice Design
ActionUS customers were notified via certified mail via USPS. OUS customers were notified through the Philips Markets Organizations. Philips is issuing two Urgent Medical Device Correction letters, one for Allura and Azurion customers and another one for MultiDiagnost Eleva Customers. Additionally, an IFU addendum will be provided to customers that includes instructions on how to handle the footswitch cable. For Allura and Azurion customers an updated preventative maintenance manual update is also being provided. Philips will contact customers to schedule a visit to inspect the foot switch and to
Quantity19,115 systems
DistributionWorldwide distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050151MULTIDIAGNOST ELEVA
Philips Medical Systems North America, Inc., Feb 9, 2005
K162859Allura Xper FD series and Allura Xper OR Table series
Philips Medical Systems Nederland B.V., Dec 2, 2016
K163715Allura Xper R9
Philips Medical Systems Nederland B.V., Jan 26, 2017
K200917Azurion R2.1
Philips Medical Systems Nederland B.V., May 1, 2020