Recall Z-2612-2023 ยท GAA Needle, Aspiration And Injection, Disposable

Tytek Medical Inc

TyTek PneumoDart- Intended for introduction into the body to facilitate removal of air from the pleural cavity in patients with a tension pneumothorax Model Number: TM-317

DateSep 19, 2023 (initiated Oct 5, 2020)
StatusTerminated
ReasonUpdated IFU to inlclude: (1) identify the risks associated with the hazard of needle decompression and (2) improve the artwork showing placement site location
Root causeDevice Design
ActionTyTek New Product IFU was initially sent out to all who purchased the product on 08Dec2020 with notification on what was updated. An official field safety notice (FSN) was sent to customers on 02Aug2021.The changes to the IFUs were proactive in nature. There has been no change to the Products themselves. When used in a manner consistent with the IFUs, the Products are safe for their intended use, and we wish to reiterate that there have been no incidents whatsoever. For you reference I have attached the IFU that we originally sent you in December 2020. Please do not hesitate to contact us shou
Quantity22,026 units
DistributionOH, SC, VA Foreign: Country Australia Denmark Finland Germany Hong Kong/China Israel SPAIN The Netherlands Turkey UAE UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.