ClearanceBrief
Recall Z-2664-2023 ยท JAK System, X-Ray, Tomography, Computed

Olympus Corporation of the Americas

Veran : Endobronchial Always on Track Models: INS-0392 AOTT 21ga Needle, 12mm L, 1.8mm OD, 5/BX INS-0382 AOTT 19ga Needle, 12mm L, 1.8 OD, 5/Bx INS-0372 AOTT Serrated Forceps, 1.8mm OD, 5/Bx INS-0362 AOTT Oval Cup Forceps, 1.8mm OD, 5/Bx INS-0352 "Always-On Tip Tracked (AOTT) Brush, 15mm L, 1.8mm OD, 5/Bx" INS-0322 View Peripheral Catheter, 3.2mm OD, 2.0mm WC INS-0305

DateSep 29, 2023 (initiated Aug 10, 2023)
StatusOpen, Classified
ReasonNon-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment
Root causeUnder Investigation by firm
ActionOlympus Corporation of the Americas ( Olympus ),[Veran Medical Technologies, a wholly owned subsidiary] issued Urgent Medical Device Recall Letter on 8/10/23 to: Endoscopy Department, Pulmonary Department, Risk Management for the VERAN ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments. Letter states reason for recall, health risk and action to take: Veran requests you to take the following actions: 1. Inspect your inventory and identify any Veran models. Please check all areas of the hospital to determine if a
Quantity60102 units
DistributionWorldwide distribution - US Nationwide and the countries of CN, DE, HK, IT, KR, SG, TH, TW.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093146IG4 IMAGE GUIDED SYSTEM, MODEL: SYS-0200
Veran Medical Technologies, Inc., Nov 30, 2009
K122106SPIN DRIVE ALWAYS-ON TIP TRACKED STEERABLE CATHETER ALWAYS-ON TIP TRACKED SHEATH ALWAYS-ON TIP TRACKED ASPIRATION NEEDLE
Veran Medical Technologies, Inc., Dec 21, 2012
K170023SPiN Thoracic Navigation System
Veran Medical Technologies, Inc., May 15, 2017