Recall Z-0154-2024 ยท NHA Abutment, Implant, Dental, Endosseous

Preat Corp

Preat REF 9007124, Neodent GM x 4mm Healing Abutment (Wide), Rx Only, Non-Sterile, Single Use Only. Dental

DateOct 24, 2023 (initiated Jul 18, 2023)
StatusOpen, Classified
ReasonThe healing abutments contain a dimensional condition that may lead to a gap between the healing abutment and the implant allowing for microleakage
Root causeProcess change control
ActionOn August 4, 2023, Preat Corporation issued a "Urgent Medical Device Recall" notification via: USPS. Preat ask consignees to take the following actions: Product within the scope of the recall shall be discarded or returned to Preat Corporation. Instructions for the return will be provided. Complete and email response form to Nichole Fetterman, E-Mail: nfetterman@younginnovations.com or mail to Preat Corporation, Attn. Nichole Fetterman, 2625 Skyway Drive, Ste. B, Santa Maria, CA 93455
Quantity4 units
DistributionUS: FL, CA, OR, TX, OH, CA, SC, OR, NJ, SC, OUS: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K220823Preat Abutments
Preat Corporation, Oct 26, 2022