ClearanceBrief
Recall Z-2566-2023 ยท IYE Accelerator, Linear, Medical

Landauer

Calibrate (Cs-137) nanoDot D2DNS, Model Numbers: a) 18125-000; b) 18125-KIT; c) 18125-SET; radiation monitoring dosimeter used with the microSTAR readers

DateSep 12, 2023 (initiated Jul 12, 2023)
StatusCompleted
ReasonLANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
Root causeUnder Investigation by firm
ActionLandauer disseminated an URGENT - Medical Device Recall notice to its consignees on 07/12/2023 by US mail and email. The notice explained the problem with the devices, potential risk, and requested the following: Discontinue use of the affected products, destroy used products, and return unused products.
Quantity13294 units
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K192196Medical Radiation Dosimetry System microSTARii
Landauer, Dec 27, 2019