ClearanceBrief
Recall Z-2482-2023 ยท MXK Device, Analysis, Anterior Segment

Oculus Optikgeraete GMBH

Oculus Pentacam AXL Wave, Ref 70020, CE 0123

DateAug 30, 2023 (initiated Aug 8, 2023)
StatusOpen, Classified
ReasonTheir is a potential that optical devices with insufficient anti-reflective coating may lead to incorrect axial length measurements.
Root causeProcess control
ActionOn August 04, 2023, Oculus issued a Urgent Field Safety Notice" Correction via UPS Express. Oculus ask consignees to take the following actions: 1. Immediately stop using the affected devices for axis length measurement (mark the affected devices with a note) or take them out of service immediately. Devices not listed in Appendix B are not affected by this corrective action and can continue to be used according to the instructions for use. 2. Please fill out the attached form (Appendix A) and return it to OCULUS and your local OCULUS representative as soon as possible, but no later than 3
Quantity143 units
DistributionUS Nationwide distribution in the states of CA, FL, MA, MN, NY, NC, OH, PA, TX, VA, WA. OUS: Not Provided
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201724Pentacam AXL Wave
Oculus Optikgerate GmbH, Oct 21, 2020