ClearanceBrief
Recall Z-2472-2023 ยท JFM Enzymatic Method, Bilirubin

Siemens Healthcare Diagnostics, Inc.

Atellica CH LDL Cholesterol- In vitro diagnostic use in the quantitative determination of LDL cholesterol in human serum and plasma (lithium heparin, EDTA, sodium heparin Test Code: LDLC Siemens Material Number (SMN): 11537214

DateAug 28, 2023 (initiated Jul 10, 2023)
StatusOpen, Classified
ReasonPotential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from Atellica CH Total Bilirubin_2 (TBil_2), Atellica CH LDL Cholesterol (LDLC), Atellica CH Gamma-Glutamyl Transferase (GGT) and Atellica CH HDL Cholesterol (HDLC)
Root causeUnder Investigation by firm
ActionSiemens Healthineers issued Urgent Medical Device Correction ACHC23-01.C.US to US affected customers via FedEx beginning on 07/10/2023; and Urgent Field Safety Notice ACHC23-01.C.OUS,OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities beginning on 07/10/2023. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Perform the instructions provided
Quantity6733 units
DistributionUS Nationwide - Worldwide Distribution Foreign: Taiwan Uruguay Vietnam
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.