Recall Z-2657-2023 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

Medtronic Neuromodulation

Restore Clinician Programmer Application Software, model number A71100, used with the Restore Neurostimulators.

DateSep 29, 2023 (initiated Jul 17, 2020)
StatusTerminated
ReasonThe original version of the A71100 Restore Clinician Programmer Application has been identified to have a compatibility issue with some legacy clinician programmer software resulting in the programmer having an inability to establish communication with the implanted neurostimulators.
Root causeSoftware design
ActionLetters dated July 2020 and flagged as "URGENT: MEDICAL DEVICE CORRECTION RESTORE CLINICIAN PROGRAMMER APPLICATION SOFTWARE UPDATE Model A71100" were issued via USPS Priority First Class Mail on 7/17/2020. The letter provided an issue summary and informed the consignee the updated version of the A71100 application was now available for download and installation for resolving the issue. The consignee was to follow the instructions provided with the letter to update the application and if assistance was needed, contact information was provided for the Technical Services team support. Once
Quantity2,408 devices
DistributionWorldwide distribution - US Nationwide (there was government and military distribution) and the countries of Australia and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.