ClearanceBrief
Recall Z-2593-2023 ยท KNN Components, Wheelchair

Aquila Corporation

APK2 Pressure-Sensing Wheelchair Cushions

DateSep 18, 2023 (initiated Jul 6, 2023)
StatusOpen, Classified
ReasonDefective battery pack in wheelchair cushions can overheat resulting in property damage and one possible injury.
Root causeNonconforming Material/Component
ActionAquila Corporation issued a "Product Safety-Corrective Action URGENT NOTICE" on 07/27/2023 via email and USPS first class mail. The notice explained the problem, the risk to the user, and to cease use of the affected product. The firm is seeking return of the products. Options also include: 1) modifying the device to run off of a powered wheelchair using instructions to be provided by the firm or 2) exchange of the batteries. Actions to be taken by the Dealer/Customer: 1. Email us at aquilarc23@gmail.com to receive your RMA#. Remove the fabric coverings from your SofTech Cushion or APK2
Quantity117 units
DistributionWorldwide distribution: US (nationwide) and OUS (foreign) countries of: Canada, Sweden, Romania, Australia, UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.