Recall Z-2522-2023 ยท FDS Gastroscope And Accessories, Flexible/Rigid

Aomori Olympus Co., Ltd.

Olympus Distal Attachment (in Olympus Disposable EMR Kit)- Designed to be attached to the distal end of the endoscope. Disposable EMR Kit: K-001, K-002, K-003, K-004, K-011, K-012

DateSep 7, 2023 (initiated Jul 27, 2023)
StatusOpen, Classified
ReasonThe Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002
Root causeNonconforming Material/Component
ActionOlympus issued Urgent Medical Device Recall Letter dated July 27, 2023 to ENDOSCOPY DEPARTMENT, RISK MANAGEMENT. Letter states reason for recall, health risk and action to take: Immediately assess any product you have to identify the product number with affected lot listed in this communication, cease use and quarantine any affected products. The image below depicts the area where the lot number is identified. The lot# is on the Carton box and Pack. Please refer to Attachment 2 for the identification of the subject products. 2. Call your Olympus customer service representative at 1-800-848-902
Quantity1,048 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K984358OLYMPUS DISTAL ATTACHMENT, MODELS MH-462,-463,-464,-465,-466,-483,-587,-588,-589,-590,-591,-592,-593,-594,-595,-596,-597
Olympus America, Inc., Sep 17, 1999