Recall Z-2442-2023 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

Boston Scientific Corporation

EMBLEM MRI S-ICD Pulse Generator Model A219; Subcutaneous Implantable Cardioverter Defibrillator

DateAug 22, 2023 (initiated Jul 11, 2023)
StatusOpen, Classified
ReasonThere is a potential for a rare interaction between the EMBLEM S-ICD and LATITUDE communicator, which may cause S-ICD sensing disablement for a 24-hour interval.
Root causeSoftware design
ActionBoston Scientific disseminated an Important Medical Device Information notice to the physician or healthcare professional consignees beginning on 07/11/2023 by hand delivery using their sales representatives, The notices explain the problem, risk, and recommend the following: Please distribute this letter to all HCPs who manage follow-ups of S-ICD patients. Perform the next scheduled device check in-person using a Model 3300 LATITUDE programmer with Model 3877 v1.04 software or Model 3200 EMBLEM programmer with Model 2877 v4.10 software to address any potential for this rare S-ICD behavior.
Quantity41,411 units
DistributionWorldwide