Recall Z-2488-2023 · CFN Method, Nephelometric, Immunoglobulins (G, A, M)

Beckman Coulter Inc.

Beckman Coulter, IVD, REF: OSR61171, IgA, 4 x 14 mL R1, 4 x 11 mL R2

DateAug 31, 2023 (initiated Jul 12, 2023)
StatusOpen, Classified
ReasonLipemic interference failed to meet the performance specification listed within the IFU.
Root causeOther
ActionOn July 18, 2023 Beckman Coulter issued a "Urgent Medical Device Recall" Notification via email and postal mail. Beckman Coulter asked consignees to take the following action: 1. No retrospective re view of past patient results is required however Beckman Coulter recommends sharing the content of this letter with your laboratory and/or medical director regarding the need to review previous patient test results. 2. Discontinuance or disposal of this product is not necessary. 3. Per the IFU, avoid highly lipemic samples when using the IgA assay. 4. No update is required to the LIH influe
Quantity12,850 units
DistributionUS: AL AR AZ CA CO CT DC DE FL GA IA IL IN KS KY LA MA MD MI MN MO MS NC ND NE NJ NM NY OH OK OR PA RI SC SD TN TX UT VA WA WI WV OUS: Algeria, Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Colombia, C¿te d'Ivoire, Croatia, Cur
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K073489OLYMPUS IGA REAGENT
Olympus America, Inc., Feb 11, 2008