Recall Z-2287-2023 ยท JEY Plate, Bone

Materialise N.V.

ProPlan CMF Patient-Specific Guide, Mandible guides (3A & 3B)-Intended to guide the marking of bone and/or guide surgical instruments in mandibular and maxillofacial surgical procedures. Identification number: SD900.101

DateAug 3, 2023 (initiated Jun 19, 2023)
StatusCompleted
ReasonWrong versions of guides 3A and 3B (Patient Specific) were shipped prior to the re-design request
Root causeProcess change control
ActionSynthes sales representative was made aware of the issue on 19 June 2023, by phone and email. Materialise requested the product to be returned on 19 June 2023, Synthes sales representative confirmed that surgery will be performed without guides and will return the case. The guides were returned/received at Materialise on 23 June 2023.
Quantity1 unit
DistributionUS Nationwide distribution in the state of KY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K103136SURGICASE CMF, SURGICASE CONNECT, SURGICASE GUIDE
Materialise NV, Mar 18, 2011