ClearanceBrief
Recall Z-2408-2023 ยท HQC Unit, Phacofragmentation

Alcon Research, LTD.

Infinity FMS Pak

DateAug 16, 2023 (initiated Jun 28, 2023)
StatusOpen, Classified
ReasonReports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.
Root causeProcess control
ActionCustomers were sent a Voluntary Medical Device Correction Notification by mail and dated 06/29/2023. Customers are asked to follow the risk mitigation precautions detailed in the notice to avoid the instance of overtightening the Phaco Tip and generating plastic particles. Additionally, customers are asked to forward the recall notification to all departments or organizations that use recalled devices. Customers are asked to complete and return the provided Response Form by fax at 817-302-4337 or email at Market.Actions@alcon.com. Customers who have experienced any adverse events or quality re
Quantity51,054 units
DistributionWorldwide distribution - US Nationwide and the countries of Aland Islands, Albania, Algeria, Andorran, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bermuda, Botswana, Brazil, Brunei Daruss., Bulgaria, Burma, Cambodia, Canada, Ca
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021566INFINITI
Alcon Laboratories, Inc., Jul 2, 2002