Recall Z-2467-2023 ยท FDT Duodenoscope And Accessories, Flexible/Rigid

Olympus Corporation of the Americas

EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V

DateAug 28, 2023 (initiated Jul 27, 2023)
StatusOpen, Classified
ReasonRecent reports of patient infection.
Root causeUnder Investigation by firm
ActionConsignees were sent an Urgent Medical Corrective Action notification dated 7/27/2023. This action is being initiated due to the firm becoming aware of and FDA inquiring about recent reports of infections and positive cultures. Olympus is reminding users to closely follow reprocessing instructions, including periodic inspections by Olympus Service, and informing users about an updated manual intended to improve comprehension of existing reprocessing steps. Additional enhancements to the manual include the use of reprocessing timelines and simplified text/visualization techniques. The manual is
Quantity6,426 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193182Evis Exera III Duodenovideoscope Olympus TJF-Q190V
Olympus Medical Systems Corp., Jan 17, 2020
K202661Evis Exera III Duodenovideoscope Olympus TJF-Q190V
Olympus Medical Systems Corp., Dec 8, 2020
K220587Evis Exera III Duodenovideoscope Olympus TJF-Q190V
Olympus Medical Systems Corp., May 11, 2022