Recall Z-2364-2023 ยท HIH Hysteroscope (And Accessories)

Scholly Fiberoptic Gmbh

UroLift Visual Obturator, 20Ch-The visual obturator is used with the UroLift system. Instrument for insertion of a cystoscope shaft into the bladder. Model: UL-VO

DateAug 9, 2023 (initiated Jun 23, 2023)
StatusOpen, Classified
ReasonThere is an improperly performed weld between the cone of the sheath lock and tube of the obturator; mechanical load capacity is impaired, and fracture of the welded joint may occur. Contamination can occur in this capillary gap, which is not covered by the reprocessing validations performed.
Root causeProcess control
ActionSCHOLLY FIBEROPTIC GMBH initiated Recall Plan Letter to Users, Importers and Distributors . The US Distributor- Teleflex (parent company of NeoTract) notified on June 23, 2023 to forward the notification letter to their customers. Letter states reason for recall, health risk and action to take: Review this Field Safety Notice in its entirety and ensure that all users of the above-mentioned product in your organization and other concerned persons are informed about this Field Safety Notice. 2. If you are a distributor, please forward this correction notification to your customers. 3. Identi
Quantity98 units
DistributionNationwide Foreign: Japan, South Korea
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150158Schoelly Cystoscopes/Hysteroscopes and Accessories
Schoelly Fiberoptic GmbH, Mar 9, 2015