Recall Z-0100-2024 ยท MSW System, Test, Thyroglobulin

Beckman Coulter, Inc.

Access Thyroglobulin Calibrators

DateOct 11, 2023 (initiated Jul 6, 2023)
StatusOpen, Classified
ReasonIt has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11% in patient results.
Root causeOther
ActionBeckman Coulter issued an Urgent Medical Device Recall letter dated 06/28/2023 to its consignees affected worldwide on 07/06/2023 via email or letter and followed with an updated letter (dated 07/07/2023) issued on 07/14/2023 which contained FDA suggested edits. The notice explained the problem, the risk to the patient and requested the following actions be taken: If you have downloaded the alternative calibrator card issued in April 2023, please discard that calibrator card, and do not use it moving forward. . Only use Access Thyroglobulin calibrator lots 338013 or higher to calibrate Acce
Quantity5980 units
Distributionworldwide except US, Canada, and China