Recall Z-2288-2023 ยท POE Extended Depth Of Focus Intraocular Lens

Johnson & Johnson Surgical Vision, Inc.

TECNIS Symfony, Extended Range of Vision IOL, REF: ZXR00, STERILE.EO, Rx Only

DateAug 3, 2023 (initiated Jul 6, 2023)
StatusOpen, Classified
ReasonTheir is a potential that intraocular lenses (IOLs) were manufactured with unintended toric fiducial marks.
Root causeUnder Investigation by firm
ActionOn June 07, 20223, Johnson & Johnson Vison issued an "Urgent: Medical Device Recall" notice to affected consignees. Consignees were asked to take the following actions: 1. Identify if any of your inventory contains TECNIS IOLs from the list on pages 5-6 (see page 4 for label examples). 2. Remove all affected TECNIS IOLs from your inventory. No other TECNIS IOLs are affected by this recall. 3. Complete the attached Customer Reply Form (on page 3). We require this information for reconciliation purposes with regulatory agencies, even if you have no inventory. 4. Follow the specific instruct
Quantity42
DistributionInternational distribution to the countries of China, India, Korea.
FDA recordOpen on fda.gov