ClearanceBrief
Recall Z-2370-2023 ยท NIE Catheter, Hemodialysis, Triple Lumen, Non-Implanted

Covidien LP

MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, IC Tray REF: 8888101004HP

DateAug 11, 2023 (initiated Jun 28, 2023)
StatusOpen, Classified
ReasonCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
Root causeProcess control
ActionMedtronic issued Urgent Medical Device Recall letter beginning on June 28, 2023. Letter states reason for recall, health risk and action to take: To help you identify if you have affected product, please visit our website www.Medtronic.com/Mahurkar-Triple-Lumen-Catheter-Recall. Here you will find a tool to help you determine if the product you have is affected by this recall. 2. Immediately quarantine and discontinue use of all unused Mahurkar TM Acute Triple Lumen Catheters and Mahurkar TM Acute High Pressure Triple Lumen Catheters referenced in Attachment B - List of affected Lot numbers (se
Quantity1325 units
DistributionUS Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Ital
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K020089MAHURKAR TRIPLE LUMEN CATHETER, 12 FR
The Kendall Company, Apr 10, 2002
K102605MAHURKAR TRIPLE LUMEN DIALYSIS CATHETER
Covidien, LLC, Formerly Registered AS Tyco Healthca, Dec 22, 2010