Recall Z-2253-2023 ยท FFL Dislodger, Stone, Basket, Ureteral, Metal

Wilson-Cook Medical Inc.

Fusion Lithotripsy Extraction Basket, Model Number FS-LXB-3X6 G48278; Fusion Lithotripsy Extraction Baskets are intended to be used for endoscopic removal of biliary stones and foreign bodies.

DateJul 28, 2023 (initiated Jun 22, 2023)
StatusOpen, Classified
ReasonNon-conforming product may appear in the field as inability to advance or retract the basket, or the drive wire breaking or separating from handle (with or without mechanical lithotripsy).
Root causeProcess control
ActionCook Medical issued an URGENT: MEDICAL DEVICE REMOVAL notice to their consignees on 07/14/2023 by email or priority mail. The notice explained the issue, the risk, and requested the consignee cease use and return the affected product.
Quantity402 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.