Recall Z-2333-2023 ยท OHD Gynecological Laparoscopic Kit
MEDLINE INDUSTRIES, LP - Northfield
Custom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. CHN OR DAVINCI PROSTATE, Model Number: DYNJ902540N; b. CUSTOM GYN, Model Number: DYNJ63257B; c. DBD-AMNIOCENTESIS PACK-LF, Model Number: DYNJ0771979C; d. DBD-GYN EXP LAPAROTOMY, Model Number: DYNJ906578F; e. DBD-GYN LITHOTOMY, Model Number: DYNJ906579F; f. DBD-GYN LS, Mo
| Date | Aug 3, 2023 (initiated May 15, 2023) |
| Status | Open, Classified |
| Reason | Potential for the labeled sterile ultrasound gel component within packs to not meet sterility specifications. |
| Root cause | Nonconforming Material/Component |
| Action | Medline notified customers beginning on 05/15/2023, via mail and email with letter. Customers with Medline Probe Cover Kits were instructed to check inventory for affected kits, destroy any affected kits, and complete and return the response form. Distributors were instructed to notify their customers and have them document and destroy affected kits. Customers with the non-sterile and sterile procedural kits were instructed to quarantine kits, complete and return the response form and, once updated labels are received, over label and discard the affected ultrasound gel product within the ki |
| Quantity | 7,462 kits |
| Distribution | Worldwide distribution - US Nationwide and the countries of Canada, China, Costa Rica, India, Lebanon, Panama, Qatar, Saudi Arabia, Singapore, United Arab Emirates. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.