Recall Z-2280-2023 ยท PSY Whole Slide Imaging System

Hamamatsu Corporation

NanoZoomer S360MD Slide scanner system-automated system for creating, viewing, and managing digital slides. The NanoZoomer System creates diagnostic-quality digital images of glass slides containing formalin-fixed paraffin-embedded (FFPE) tissue. Catalog Number: C13220-01MD

DateAug 1, 2023 (initiated Jun 6, 2023)
StatusOpen, Classified
ReasonIf the wrong barcode information is recorded in the NDPi file, a pathologist may refer to the whole slide image (WSI) of a different patient.
Root causeSoftware design
ActionHamamatsu Corporation issued Urgent: Medical Device Correction Letter via Email on 6/6/23. Letter states reason for recall, health risk and action to take: The risk of recurrence will be fully eliminated by conducting the NZAcquire software update on the user PCs either by a Field Service Engineer being on site or via a remote control. As a precautionary measure, there are two kinds of risk mitigations in place to reduce the potential serious risk. 1. There is a clear note in the NZAcquireMD Reference manual regarding the barcode information. 2. There is a warning in the system manual that th
Quantity1 unit
DistributionNY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213883NanoZoomer S360MD Slide scanner system
Hamamatsu Photonics K.K., Sep 27, 2022