Recall Z-2255-2023 ยท PPM General Purpose Reagent

NeuMoDx Molecular Inc

NeuMoDx Cartridge-used for extraction, purification, amplification and detection of nucleic acids on the NeuMoDx 288 and NeuMoDx 96 Molecular Systems (NeuMoDx System(s) Ref: 100100

DateJul 28, 2023 (initiated May 24, 2023)
StatusOpen, Classified
ReasonTop label on some of the NeuMoDx cartridges within these lots may have been incorrectly applied during the manufacturing process, resulting in encroachment into the PCR region. The misplaced label may interfere with the module s optics performance, thereby increasing the risk for false positive results
Root causeEmployee error
ActionQiagen issued customers "Urgent: Medical Device Correction" Letter via email on 5/24/23. Letter states reason for recall, health risk and action to take: If you have remaining stock of cartridges LOT 117514, 117524 or 117528, REF 100100, do not use it. Please contact QIAGEN Technical Service for a free-of-charge replacement. Dispose of the product LOT 117514, 117254, and 117528 in accordance with your national and local safety and environmental regulations. If you have already used NeuMoDx Cartridges from this LOT, please review all results obtained with the laboratory director and assess wh
Quantity20,016 cartridges
DistributionUS Nationwide Distribution to states of: Kansas, Florida, Texas, Ohio, New York, Michigan, Pennsylvania; and OUS (Foreign) to countries of: AT, BE, CH, DE, ES, FI, FR, GB, IT, LV, PL, RO, SE