Recall Z-2237-2023 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter, Inc.

Ringed DxI Reaction Vessels (RVs)

DateJul 26, 2023 (initiated Jun 20, 2023)
StatusOpen, Classified
ReasonBeckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an internally used service test to evaluate washing efficiency. Increased relative light units (RLU) suspected to be caused by the resulting rough interior surface may lead to artificially falsely elevated results for susceptible tests, in particular, sandwich assays including, but not limited to, hsTnI, Hybritech p2PSA, IPTH/IPTH IO.
Root causeUnder Investigation by firm
ActionA customer letter titled, Urgent Medical Device Recall, dated 06/07/2023, was issued to the impacted consignee via first-class mail and email. The consignee is instructed to discard the product and request a replacement. The consignee is asked to complete the response form and submit to the recalling firm.
Quantity213 cases of 10,000 RVs (2,130,000 RVs)
DistributionWorldwide - US Nationwide distribution in the states of CA, MA, TX, NJ, MI, GA, NC, FL, WV and the country of Qatar.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023764UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM, MODEL 973100
Beckman Coulter, Inc., Jan 28, 2003