ClearanceBrief
Recall Z-2104-2023 ยท FRN Pump, Infusion

Baxter Healthcare Corporation

The Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009

DateJul 13, 2023 (initiated Jun 15, 2023)
StatusOpen, Classified
ReasonA medical device correction is being issued for SIGMA Spectrum Infusion System (V8 Platform) and Spectrum IQ Infusion System with Dose IQ Safety Software infusion pumps due to an increase in reported false upstream occlusion alarms following upgrades to software versions v8.01.01 and v9.02.01.
Root causeUnder Investigation by firm
ActionCustomers were contacted via phone and email beginning on June 2, 2023. An Urgent Medical Device communication was sent to all affected customers on June 15, 2023, via U.S.P.S., first class mail. The notice explained the problem, the risk involved, and requested the following: "Actions to be taken by Customers: 1. Operators may continue to use Spectrum V8 and Spectrum IQ pumps by following on-screen instructions, or the infusion setup instructions in the Preparing the Pump and IV Sets and Programming the Pump sections and upstream occlusion alarm troubleshooting in the Alarms section of t
Quantity19861 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K230041Spectrum IQ Infusion System with Dose IQ Safety Software
Baxter Healthcare Corporation, Mar 31, 2023