ClearanceBrief
Recall Z-2419-2023 ยท GCJ Laparoscope, General & Plastic Surgery

Carefusion 2200 Inc

Snowden-Pencer MIS DIAMOND-TOUCH CLAMP DEBAKEY 25 MM JAW 5MM, 32CM, RATCHETED, REF SP90-7043; endoscopic surgical device

DateSep 30, 2024 (initiated Jul 13, 2023)
StatusOpen, Classified
ReasonSpecific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process.
Root causeNonconforming Material/Component
ActionBD issued an URGENT: Medical Device Product Recall notice to its consignees on 13 July 2023 via FedEx 2-day and email. The notice explained the problem with the device, the risk to the patient, and requested the following actions: Medical facilities: - Discontinue the use of the affected device and destroy it - Share the notice with all users within your facility network - If the product was purchased through a distributor, contact the distributor directly for further instructions. Distributors were directed to Identify all customers within your distribution network that purchased any aff
Quantity8 units
DistributionUS and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K930667REUSABLE LAPAROSCOPIC INSTRUMENTS
Snowden-Pencer, May 16, 1994