Recall Z-2282-2023 ยท EOX Esophagoscope (Flexible Or Rigid)

Covidien, LLC

Covidien Cytosponge Cell Collection Device, REF: CYTO-201; and Covidien Cytosponge Cell Collection Kit, REF: CYTO-KIT-USM

DateAug 2, 2023 (initiated Jun 2, 2023)
StatusOpen, Classified
ReasonCell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which could lead to device fragments in patient, obstruction, airway obstruction, secondary intervention, secondary intervention (with the primary procedure), supraglottic airway obstruction, and aspiration.
Root causeProcess control
ActionOn 6/2/23, Medtronic sent recall notices to customers who were informed of the following: 1) If product was purchased directly from Medtronic, return affected product to Medtronic. 2) If product was purchased from a distributor, return affected product to the distributor. 3) The recall notice should be passed on to all who need to be aware within your organization or to any organization where the potentially affected product has been transferred. 4) Complete and return the Customer Confirmation Form to rs.gmbfcamitg@medtronic.com Adverse reactions or quality problems experienced with this pro
Quantity6902
DistributionWorldwide - US Nationwide distribution in the states of NY, MN, VA, WI and the countries of Canada, Netherlands, Poland, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181020Cytosponge Cell Collection Device
Covidien, LLC, Aug 16, 2018