ClearanceBrief
Recall Z-2184-2023 ยท MCY Wound Dressing Kit

MEDLINE INDUSTRIES, LP - Northfield

Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) LACERATION TRAY, Model Number DYNDL1134A; b) ULTRASOUND GUIDED IV/MIDLINE K, Model Number DYNDV2385; c) PEDIATRIC CENTRAL LINE KIT, Model Number DYNJ38885B; d) CENTRAL LINE TRAY, Model Number DYNJ42902B; e) CENTRAL LINE PACK - COMPANION, Model Number DYNJ47215D; f) PICC LINEPACK (PCLUI)642-LF, Model Num

DateJul 20, 2023 (initiated May 18, 2023)
StatusOpen, Classified
ReasonProbe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
Root causeDevice Design
ActionMedline distributed their recall notices to their consignees beginning on 05/18/2023 by letter. The notices explained the problem, and the risk and requested the destruction of the products. Distributors were directed to notify their consignees, have the customers document and destroy any affected product, and include your customers quantities on the response form. Beginning on 05/23/2023, Medline issued another notice to a separate group of consignees requesting that the kits be over labeled. Distributors were directed to notify the customers and have them return any affected product to
Quantity11413 units
DistributionWorldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.