Recall Z-2167-2023 ยท OFF Central Venous Catheter Tray
MEDLINE INDUSTRIES, LP - Northfield
Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ARTERIAL BUNDLE NO CATHETER, Model Number ART995; b) ART LINE KIT NO CATH, Model Number ARTNC145; c) CENTRAL LINE TRAY, Model Number CVI4705; d) CENTRAL LINE INSERTION PACK, Model Number CVI4805; e) ED CENTRAL LINE KIT, Model Number CVI4865; f) NO CATH MIDLINE INSERTION KIT, Model Number CVI4950A; g) CE
| Date | Jul 20, 2023 (initiated May 18, 2023) |
| Status | Open, Classified |
| Reason | Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams. |
| Root cause | Device Design |
| Action | Medline distributed their recall notices to their consignees beginning on 05/18/2023 by letter. The notices explained the problem, and the risk and requested the destruction of the products. Distributors were directed to notify their consignees, have the customers document and destroy any affected product, and include your customers quantities on the response form. Beginning on 05/23/2023, Medline issued another notice to a separate group of consignees requesting that the kits be over labeled. Distributors were directed to notify the customers and have them return any affected product to |
| Quantity | 269985 units |
| Distribution | Worldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.