Recall Z-2158-2023 ยท OGJ Nerve Block Tray

MEDLINE INDUSTRIES, LP - Northfield

Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) NERVE BLOCK ACCESSORY PACK, Model Number DYNJRA1220; b) PERI NEURO BLOCK PACK-LF, Model Number DYNJRA1773; c) PERI NEURO BLOCK PACK-LF, Model Number DYNJ51107B; d) PERI NEURO BLOCK PACK-LF, Model Number DYNJ0703127A; e) PERI NEURO BLOCK PACK-LF, Model Number DYNJ42367B; f) PERI NEURO BLOCK PACK-LF, Model

DateJul 20, 2023 (initiated May 18, 2023)
StatusOpen, Classified
ReasonProbe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
Root causeDevice Design
ActionMedline distributed their recall notices to their consignees beginning on 05/18/2023 by letter. The notices explained the problem, and the risk and requested the destruction of the products. Distributors were directed to notify their consignees, have the customers document and destroy any affected product, and include your customers quantities on the response form. Beginning on 05/23/2023, Medline issued another notice to a separate group of consignees requesting that the kits be over labeled. Distributors were directed to notify the customers and have them return any affected product to
Quantity23938 units
DistributionWorldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia.