Recall Z-2478-2023 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Siemens Healthcare Diagnostics Inc

epoc BGEM Test Card (BUN & TCO2),Veterinary with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1

DateAug 28, 2023 (initiated Jun 1, 2023)
StatusOpen, Classified
ReasonThere is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
Root causeUnder Investigation by firm
ActionConsignees were notified via an Urgent Medical Device Correction letter beginning June 5, 2023. Consignees were informed of the issue and were provided a workaround and actions to take. The next sensor configuration/software version, sensor configuration 42.1/epoc Host SW v3.38.2 /epoc NXS SW v4.11.11, is available to customers in Siemens Healthineers Document Library or the epoc Live Update Service (eLUS). This next sensor configuration, 42.1, mitigates the issue.
Quantity348 ea
DistributionUS distribution to AL, AR, AZ, CA, CO, DC, DE, FL, IA, ID, IL, KS, KY, LA, MI, MN, MO, NC, ND, NH, NY, OH, OK, SC, SD, and TX OUS distribution to Canada, China, Colombia, Finland, India, Mexico, Norway, Sweden, United Arab Emirates, and Vietnam
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200107epoc Blood Analysis System
Epocal, Inc., Jun 5, 2020