Recall Z-2222-2023 ยท DSP System, Balloon, Intra-Aortic And Control

Datascope Corp.

SENSATION PLUS 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number SENSATION PLUS 8Fr. 50cc IAB WITH ACCESSORIES D684-00-0576-01 SENSATION PLUS 8Fr. 50cc IAB WITH ACCESSORIES (WITH PRESSURE D684-00-0576-03 SENSATION

DateJul 21, 2023 (initiated Jun 23, 2023)
StatusOpen, Classified
ReasonDuring IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
Root causeUnder Investigation by firm
ActionFirm notified customers on June 23, 2023 via letter. Customers were made aware of the potential for the device to break during use and the risk of using the device. Customers were given alternate solutions to continue using the IAB Catheter with alternative sheath/dilator combinations. Customers may also return their affected product for a credit. The firm will continue to ship product to customers who choose to continue receiving it. However, product currently being shipped is potentially affected by the same issue. Datascope/Getinge has identified the issue with the introducer dilator and
Quantity321,609 total kits
DistributionDomestic distribution nationwide. International distribution worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133074MEGA 8FR. 50CC AND SENSATION PLUS 8FR. 50CC INTRA-AORTIC BALLOON CATHETERS AND ACCESSORIES