Recall Z-1971-2023 ยท HQF Laser, Ophthalmic

Ellex Medical Pty Ltd.

Integre Pro Yellow LP561, Integre Pro Yellow LP6RY, Integre Pro Scan Green LP6G, Integre Pro Green LP5532 Spare, Optics Bench V2, L2G/LP6G

DateJun 27, 2023 (initiated Mar 16, 2023)
StatusOpen, Classified
ReasonWhen using a Laser-Indirect-Ophthalmoscope (LIO) as the delivery device, a scattered laser beam inside the laser cavity system may unintentionally exit through the Slit Lamp output port (SDS) and get emitted from the Objective Lens when the laser is fired.
Root causeRadiation Control for Health and Safety Act
ActionEllex Safety Bulletin is the customer notification letter. Ellex will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. For questions contact your local Ellex-Lumibird representative or email us on eservice@ellex.com.
Quantity57 systems in total (6 units shipped to United States)
DistributionUS nationwide/ Worldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081565INTEGRE PRO, MODEL L2RY
Ellex Medical Pty, Ltd., Jul 2, 2008
K142398Integre Pro Scan Green, Integre Pro Scan Yellow, Integre Pro Scan Red-Green, Integre Pro Scan Red-Yellow
Ellex Medical Pty, Ltd., Jul 14, 2015