Recall Z-2226-2023 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

Medtronic Neuromodulation

A901 Communication Manager Application, used in conjunction with the A710 Clinician Programmer application..

DateJul 21, 2023 (initiated Jun 21, 2023)
StatusOpen, Classified
ReasonWhen an update to the A710 Clinician Programmer software application for the Intellis Neurostimulation System was released, the corresponding version of the A901 Communication Manager software application was not available for download.
Root causeUnder Investigation by firm
ActionOnly one of the two consignees received the notification since one of the consignees independently updated their tablet. On 6/22/2023, the notification dated June 2023 was issued to the consignee via visit by the sales representative. The notification described the issue and provided required actions which included: (1) Update the A901 Comm Manager application to version 1.0.1239 on the clinician programmer (tablet), serial #NPL3011706, using the instructions listed in Appendix A, Updating Clinician Programmer Software; (2) Complete the Customer Confirmation Form enclosed with the letter ac
Quantity2 clinician tablets
DistributionInternational distribution to the countries of Israel and Serbia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.