Recall Z-2208-2023 ยท DPW Flowmeter, Blood, Cardiovascular

Baxter Healthcare Corporation

GEM FLOW COUPLER Monitor, PN 5156-00000-011

DateJul 20, 2023 (initiated Jun 19, 2023)
StatusOpen, Classified
ReasonAn issue was identified where the LCD monitor display may not indicate the correct active channel. The issue occurs when the monitor is turned on using battery power, and channel B was selected as the last active channel prior to powering down.
Root causeUnder Investigation by firm
ActionSynovis Micro Companies Alliance (MCA), Inc., a subsidiary of Baxter International Inc., is issuing an Urgent Medical Device Correction to its consignees on 06/19/2023 by USPS First Class Mail. The notice explained the problem with the device, risk, and provided the following method to circumvent the problem: "The affected flow coupler monitors may continue to be used. To avoid encountering this issue, please use the monitor while connected to an AC power supply whenever possible. However, if using battery power, Baxter advises users to select channel A prior to shutting down the monitor. U
Quantity122 units
DistributionUS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191252Gem FlowCoupler System
Baxter/ Synovis Micro Companies Alliance, Inc., Jul 9, 2019